EVALUASI PERFORMA MIKROTOMOGRAFI SKY SCAN 1173 UNTUK MATERIAL DENSITAS RENDAH

PERFORMANCE ANALYSIS OF MICROTOMOGRAPHYSKYSCAN 1173 FOR LOW DENSITY MATERIALS

Undergraduate Theses from JBPTITBPP / 2017-09-27 14:40:51
Oleh : WA ODE SRIWAYU (NIM : 20210001); Pembimbing : Dr. rer.nat. Freddy Haryanto., S1 - Department of Physics
Dibuat : 2012, dengan 7 file

Keyword : micro ct scan, phantom, CT Analyser, performa micro ct- SkyScan 1173.

Evaluasi performa dalam menentukan kualitas citra menjadi suatu hal yang lazim dilakukan sejak micro ct scan menjadi alat yang rutin digunakan dalam studi pre klinis atau small animal imaging. Sistem mikrotomografi SkyScan1173 tipe energi tinggi dimiliki oleh Laboratorium BSC-ITB dengan spesifikasi < 130kV (Hamamatsu L9181-02) mempunyai kemampuan dalam mencitrakan material dengan kontras densitas yang tinggi. Oleh karena itu, sebuah phantom didesain dan difabrikasi menyerupai densitas jaringan tubuh manusia dengan mengacu pada standar yang dikeluarkan oleh AAPM untuk menguji kemampuan mico ct-SkyScan 1173 dalam menghasilkan citra terhadap material yang mempunyai










































kontras densitas rendah. Performa micro ct dievaluasi terhadap perubahan parameter tegangan, eksposur, dan resolusi kamera. Analisis kuantitatif dilakukan menggunakan perangkat lunak CTAn dan ImageJ untuk menentukan skala kontras, uniformity, SNR, dan low contrast criteria/CNR. Hasil yang diperoleh menunjukkan bahwa pada tegangan 30 kV dengan perubahan eksposur 0.155










































mAs, 0,160 mAs, dan 0,165 mAs, serta aplikasi resolusi kamera 560x560 piksel dan 1120x1120 piksel memberikan nilai skala kontras, uniformity, SNR, dan low contrast criteria/CNR yang signifikan untuk material yang mempunyai kontras densitas rendah

Deskripsi Alternatif :

Performance evaluation in determining the image quality be a commonly thing to do since the micro-ct scan became a routine tools used in pre-clinical studies or small animal imaging studies. Microtomography system- SkyScan1173 is the










































high-energy type of micro ct that owned by BSC-ITB Laboratory. It has voltage < 130kV (Hamamatsu L9181-02) specifications with the capability for imaging materials with high density contrast. Therefore, a phantom was designed and fabricated mimics the density of human tissue with reference to the standards issued by the AAPM. Phantom used for evaluation the ability of micro-ct










































SkyScan 1173 in imaging materials which having a low density contrast. Performance evaluation of micro ct is done by changing the voltage, exposure, and camera resolutions parameters. Quantitative analysis is performed using ImageJ and CTAn software to determine the contrast scale, uniformity, SNR, and low contrast criteria/CNR. The results was obtained show that a voltage of 30 kV










































by changing the exposure at 0.155 mAs, 0.160 mAs, and 0.165 mAs, as well as the application of camera resolution 560x560 pixels and 1120x1120 pixels giving a significant values of contrast scale, uniformity, SNR, and low contrast










































criteria/CNR for imaging materials which have low density contrast.

Beri Komentar ?#(0) | Bookmark

PropertiNilai Properti
ID PublisherJBPTITBPP
OrganisasiS
Nama KontakUPT Perpustakaan ITB
AlamatJl. Ganesha 10
KotaBandung
DaerahJawa Barat
NegaraIndonesia
Telepon62-22-2509118, 2500089
Fax62-22-2500089
E-mail Administratordigilib@lib.itb.ac.id
E-mail CKOinfo@lib.itb.ac.id

Print ...

Kontributor...

  • Pembimbing : Dr. rer.nat. Freddy Haryanto, Editor: PKL-SMK

File PDF...