PENGEMBANGAN DAN REALISASI STETOSKOP ELEKTRONIK BERBASIS PC UNTUK AUSKULTASI JANTUNG DAN PERNAFASAN

Development and Realization of PC Based Electronic Stethoscope for Heart and Respiratory Auscultation

Undergraduate Theses from JBPTITBPP / 2017-09-27 10:18:41
Oleh : PRIMUS PANDUMUDITA (NIM 13201138), S1 - Department of Electrical Engineering
Dibuat : 2007, dengan 7 file

Keyword : Stetoskop, Sinyal Suara Jantung, Sinyal Suara Pernafasan, Pengolahan Sinyal, MATLAB

Dalam laporan tugas akhir ini dibahas tentang perancangan alat yang merupakan pengembangan dari stetoskop konvensional. Fungsi utama yang dikembangkan dalam alat ini adalah untuk mendengarkan suara yang dihasilkan oleh organ tubuh, layaknya stetoskop konvensional. Kelebihan pada alat ini adalah adanya sistem pengolahan sinyal untuk menghasilkan keluaran yang lebih baik, juga kemampuan untuk menyimpan data auskultasi.

Kemampuan auskultasi dengan menggunakan stetoskop konvensional bergantung pada kemampuan pendengaran atau pengalaman pengguna. Hal ini sering menimbulkan perbedaan analisis dari setiap pengguna. Permasalahan ini diharapkan dapat diperbaiki dengan adanya sistem pemrosesan sinyal dan gambar tampilan sinyal sebagai referensi alternatif selain suara. Dalam realisasi alat dapat dibagi menjadi tiga bagian: penerima sinyal, perangkat keras dan perangkat lunak.

Penerima sinyal menggunakan kepala stetoskop dan electret condensor microphone yang berfungsi untuk menangkap sinyal suara aktifitas jantung dan sinyal suara pernafasan. Sinyal suara auskultasi tersebut kemudian diolah menjadi sinyal listrik oleh perangkat keras agar dapat menjadi masukkan bagi perangkat lunak pada PC. Bagian pada perangkat keras terdiri rangkaian penguat dan filter menggunakan OpAmp 741, serta rangkaian penguat daya menggunakan LM386. Catu daya yang digunakan perangkat keras adalah 2 buah batere 9 volt. Pemilihan batere sebagai catu daya agar ruang gerak penggunaan alat tidak terbatas.

Perangkat lunak yang dikembangkan berfungsi untuk mengolah sinyal keluaran dari perangkat keras. Pengolahan sinyal yang dilakukan perangkat lunak adalah pemisahan sinyal suara jantung dan pernafasan, melakukan penghitungan dan analisa terhadap detak jantung, serta menghasilkan keluaran audio dan visual dari sinyal suara jantung dan sinyal suara pernafasan. Perangkat lunak tersebut juga memiliki kemampuan untuk menyimpan data auskultasi beserta data pasien yang memungkinkan data auskultasi untuk dibuka dan dianalisa kapan saja. Pengembangan perangkat lunak dalam PC menggunakan MATLAB.

Deskripsi Alternatif :

This final project is about development of a device that principally adopted from conventional stethoscope. Main basic function of the device being developed is to hear the audio signal that is generated by human tissue, similar with conventional stethoscope. The difference is this device has a signal processing system to produce better output, also ability to store auscultation data therefore can be opened for the next examination or to be compared with the other data. The signal processing system includes regulation of frequency range that is adjusted with the kind of input signal.


With the additional feature that is included in this device, it is expected it can give solution of several problems in using conventional stethoscope. The conventional stethoscope ability in data acquisition is differ depend on the hear ability or clinical experience of the user, and cause the appearance of different analysis from each user. This problem could be fixed by the signal processing system and signal display for another reference beside the sound. In the realization of the device, it's divided three parts: signal receiver, hardware, and software.

Signal receiver is using stethoscope head and electret condensor microphone to capture heart activities and breathing sound signals. These auscultation signals were transformed into electricity signal by hardware, to become an input for PC's software. Hardware parts consist of amplifier circuit and filter circuit using OpAmp 741, and power amplifier circuit using LM386. Power supply of the hardware is two 9 volt batteries. Portability of the device can be maintained using this kind of power supply.

The software is developed in PC using MATLAB. The software has function to process output signal from hardware. The signal processing held in software are separating heart and breathing sound signal, counting and analyzing heartbeat, and also produce audio and visual output of the heart and breathing sound signals. Besides signal processing, this software has the ability of auscultation data and patient data storing. Which enable user to open and analyze the auscultation data anytime.

Copyrights : Copyright Â(c) 2001 by ITB Central Library. Verbatim copying and distribution of this entire article is permitted by author in any medium, provided this notice is preserved.

Beri Komentar ?#(0) | Bookmark

PropertiNilai Properti
ID PublisherJBPTITBPP
OrganisasiS
Nama KontakUPT Perpustakaan ITB
AlamatJl. Ganesha 10
KotaBandung
DaerahJawa Barat
NegaraIndonesia
Telepon62-22-2509118, 2500089
Fax62-22-2500089
E-mail Administratordigilib@lib.itb.ac.id
E-mail CKOinfo@lib.itb.ac.id

Print ...

Kontributor...

  • Pembimbing: Prof. Dr. Ir. Soegijardjo Soegijoko, Editor: Vika A. Kovariansi

File PDF...